Brand: STIK
Model: ILW-185TI
I. Product technical indicators:
1. Main use
It is used for the isolation and cultivation of the new coronavirus required for the prevention and control of the epidemic in the General Administration, and provides sample resources for the whole genome sequencing of a small number of samples such as the new coronavirus, as well as the sequencing analysis of unknown pathogens.
2. Special qualification requirements: None.
3. Main technical parameters.
3.1 Working volume: 185L, built-in double high-efficiency circulation filtration system (HEPA).
3.2 Carbon dioxide concentration control system: high-precision infrared (IR) sensor.
3.3 CO2 control range: 0~20%, CO2 control accuracy: ±0.1%.
3.4 Working environment temperature: 5-50℃.
3.5 Temperature control range: 5℃~60℃ higher than room temperature, temperature control accuracy: ±0.1℃, temperature uniformity: ±0.4℃ (at 37℃).
3.6 Working environment humidity: 20-80%.
3.7 Power supply: 220V±10%, 50 Hz±1.
3.8 Sterilization function: alcohol disinfection wipe.
3.9 Air filtration system: HEPA high-efficiency filtration system, air cleanliness: Class 100 (within 5 minutes).
3.10 Safety control system: temperature alarm, CO2 concentration alarm, power failure alarm, door opening alarm.
3.11 With program self-check function.
3.12 With gas automatic control system.
3.13 The glass inner door has a heating function to prevent condensation.
3.14 Optional automatic cylinder switching device, which can automatically judge and automatically switch cylinders when the gas is used up.
3.15 Shelves are standard with 4 pieces.
3.16 With HEPA air filtration system, the chamber air is filtered every 60s.
3.17 There is a microbial filter at the air inlet.
3.18 Built-in probes and sensors are close to cell cultures.
3.19 Single box, stackable.
3.20 Comes with a touch screen to provide full data visibility.
II. Configuration list:
1. CO2 incubator 1.
2. 1 set of connecting pipes.
3. 4 baffles, 1 HEPA filter, 1 water tray.